Protector XStream I-S Laboratory Hoods
The Protector XStream I-S Laboratory Hood takes containment, safety and energy savings to the extreme. With integrated sash position blower controls, this hood was designed for use with Intelli-Sense Multi-Speed Blowers. As the sash is raised or lowered, sensors alert the blower to change speeds to maintain a safe face velocity. That face velocity is displayed on the factory-installed Guardian™ Digital Airflow Monitor for users’ peace of mind.
This general chemical, high performance fume hood was independently tested and challenged well beyond the SEFA-1 standards. With a face velocity of 40 fpm and sash fully open, the Protector XStream I-S was subjected to 50 fpm cross drafts, NIH protocol, and tracer gas measurements in the chest of the mannequin. In all scenarios, the Protector XStream I-S allowed 0.00 ppm average level of tracer gas outside the fume hood. Although your safety officer will determine the actual face velocity setting for your laboratory, the ability of this hood to contain under these adverse conditions sets a new standard of safety.
Safety is foremost, but energy savings is equally impressive. Although face velocity is a factor, it’s the volumetric rate (CFM) that determines the energy consumption. Operating a 6′ Protector XStream I-S at 60 fpm face velocity, with the sash in its fully open position, requires only 690 CFM. Regardless of your desired operating face velocity, the Protector XStream I-S yields the lowest required CFM of any hood.
Our most energy-efficient benchtop hoods have containment-enhancing features that allow them to operate at face velocities as low as 60 fpm. Their integrated sash position sensors automatically adjust the blower speed as the sash is raised or lowered.
Features (All Models)
By-pass airflow design
Eco-Foil™ Air Foil with aerodynamic Clean-Sweep™ airflow openings
Cord-Keeper™ Slots on left and right side of air foil
Upper Dilution Air Supply
Factory-installed Intelli-Sense Blower control box
Factory-installed sash position activated blower speed switches
Factory-installed Guardian™ Digital Airflow Monitor
Glacier white powder-coated steel exterior
Rear Downflow Dual Baffle System
Chemical-resistant, fiberglass-reinforced, composite panel liner and baffles with flame spread index less than 25 per ASTM E84. Baffles are removable for cleaning
Opti-Zone™ Baffle with tapered slots
Tempered safety glass vertical-rising sash
Powder-coated sash handle with aerodynamic Clean-Sweep™ openings
37.5″ (95.3 cm) high sightline from the work surface to the header panel
Removable front and side panels, and front and interior service access panels for access to plumbing and electrical wiring
Pre-wired LED lighting with vapor-proof design and ADA-compliant light and blower switches. Lighting rated to 50,000 hours.
Sash stop located at 18″ (45.7 cm) sash opening position
Powder-coated stainless steel, 12.81″ (32.5 cm) ID exhaust connection(s)
Fixtured Models May Feature
Two pre-plumbed service fixtures with forged brass valves, lower right side with brass tubing for gas and lower left side with copper tubing for cold water. Components for converting either or both fixtures to air and vacuum are provided. Inlet tubing is not provided.
One pre-wired GFCI electrical duplex receptacle on lower right side and, on 8′ models only, one additional pre-wired GFCI electrical duplex receptacle on lower left side
Required Accessories (Not Included)
Intelli-Sense Multi-Speed Remote Blower (without control box)
Ductwork
Work Surface
Base Cabinet or Stand
Dedicated Exhaust
Optional Accessories
Service Fixture Kits (Some models include service fixtures)
Electrical Duplex Kits (Some models include electrical duplex)
Distillation Grid Kits
Sash Stop Kits
Ceiling Enclosure and Rear Finish Panel Kits
Compliance
CFR 29, Part 1910
Exceed SEFA 1 definition when tested to ASHRAE 110
NFPA 45
ASTM E84-09C
ANSI Z9.5
UL 61010-1
CAN/CSA C22.2 No. 61010-1
UL 1805
CE Conformity Marking (230 volt models)
SEFA 8
ADA (Contact Labconco for ADA-compliant models)